Getting Started and Checkout

This chapter describes the check out of the Agilent 1200 Series fluorescence
detector using the Agilent isocratic checkout sample.
When required
Hardware required
Parts required

Starting Your Detector

1 Turn ON the detector.
2 Turn ON the lamp.
3 You are now ready to change the settings of your detector.
1200 Series FLD User Manual
If you want to checkout the detector
LC system with G1321A FLD
Start-up Kit 5063-6528, includes
LC cartridge Hypersil ODS, 5um, 125x4mm with CIS cartridge holder
Agilent isocratic checkout sample (Part number 01080-68704)
Fittings, Qty=2 (Part number 0100-1516)
Capillary, 150 mm long, 0.17 mm i.d. (Part number 5021-1817)
When the lamp is turned on the first time the instrument performs some
internal checks and a calibration check which takes about 5 minutes.