Lamp Intensity Test - Agilent Technologies 1200 Series User Manual

Fluorescence detector
Hide thumbs Also See for 1200 Series:
Table of Contents

Advertisement

Lamp Intensity Test

When required
Tools required
Pre-requisites
The intensity test scans an intensity spectrum via the reference diode
(200 - 1200 nm in 1 nm steps) and stores it in a diagnosis buffer. The scan is
displayed in a graphic window. There is no further evaluation of the test.
Results of this test are stored as lamp history (date code, intensity).
Figure 55
1200 Series FLD User Manual
If the flow cell or lamp has been replaced
None
clean flow cell (flushed)
Lamp Intensity Test (Report)
Maintenance
8
123