Setting The Chromatographic Conditions - Agilent Technologies 1200 Series User Manual

Fluorescence detector
Hide thumbs Also See for 1200 Series:
Table of Contents

Advertisement

4
First Steps with the Detector

Setting the Chromatographic Conditions

1 Set up the system with the following chromatographic conditions and wait
Table 6
Mobile phases
Column
Sample
Flow rate
Compressibility A (water)
Compressibility B (Acetonitrile)
Stroke A and B
Stop time
Injection volume
Oven temperature (1200)
FLD Excitations/Emission Wavelength
FLD PMT Gain
FLD Response time
60
until the baseline gets stable.
Chromatographic Conditions
A = water = 35%
B = Acetonitrile = 65%
OSD-Hypersil column, 125 mm x 4 mm i.d. with 5
µm particles
Isocratic standard sample, 1:10 diluted in
methanol
1.5 ml/min
46
115
auto
4 minutes
5 µl
30°C
EX = 246 nm, EM = 317 nm
PMT = 10
4 seconds
1200 Series FLD User Manual