Information Messages - Kodak RVG 6500 IPS System User's Installation Manual

Wireless digital intra-oral x-ray system
Hide thumbs Also See for RVG 6500 IPS System:
Table of Contents

Advertisement

Information Messages

An information "I" error code with a message is displayed on the popup on your computer screen.
The following table lists the information messages and the action to take.
Table 3
Information Messages
Information
Error Code
Communication problem with the wireless RVG
sensor. Turn off then on your sensor or restart
I 1
your computer. If the problem persists, contact
your representative.
Wireless sensor error. Turn off then on your
I 2
sensor and restart the acquisition software. If
the problem persists, contact your
representative.
Error during image acquisition. Turn off then on
I 3
your wireless sensor and restart the acquisition
software. If the problem persists, contact your
representative.
Memory allocation problem. Turn off then on your
I 4
wireless sensor. If the problem persists, contact
your representative.
I 5
Wireless sensor has storage/transmission
memory problem. Contact your representative.
I 6
Incorrect IPS calibration. Contact your
representative
I 7
IPS Batteries have run down. Change the
batteries.
Writing transmission error. Turn off then on your
I 9
wireless sensor and restart the process. If the
problem persists, contact your representative.
Reading transmission error. Turn off then on your
I 10
wireless sensor and restart the process. If the
problem persists, contact your representative.
Information Message
KODAK RVG 6500 System_User & Installation Guide (SM728)_Ed01
Action
Turn Off then ON the RVG sensor.
Or reboot your computer.
If the problem persists, contact your
representative.
Turn Off then ON the RVG sensor.
1.
to access the RVG
Click
2.
Acquisition interface .
If the problem persists, contact your
representative.
Turn Off then ON the RVG sensor.
1.
Click
to access the RVG
2.
Acquisition interface .
If the problem persists, contact your
representative.
Turn Off then ON the RVG sensor.
If the problem persists, contact your
representative.
Contact your representative.
Contact your representative.
Change the batteries of the IPS aiming
ring.
Turn Off then ON the RVG sensor.
1.
Redo the process
2.
If the problem persists, contact your
representative.
1.
Turn Off then ON the RVG sensor.
Redo the process
2.
If the problem persists, contact your
representative.
61