Launching the X-Ray

To launch the x-ray, follow these steps:
Ask the patient to remain still.
1
Position yourself either 2 meters behind the x-ray generator or outside the door.
2
Keep visual contact with the patient during the x-ray.
3
Trigger the x-ray with the remote control of the x-ray generator.
4
The image appears in the preview screen of the RVG Acquisition interface. The light on the remote
control button blinks blue indicating the image transmission.
When the acquisition ends, the
displayed in the Imaging Window. The light on the remote control button becomes solid blue.
Check the image quality. If not satisfactory, redo the x-ray.
5
If satisfactory, remove the generator tube head.
6
Remove the RVG sensor from the patient's mouth. Remove the hygienic sensor protection.
7
Important: DO NOT pull the sensor by its cable when you remove
the hygienic protection.
RVG Acquisition interface disappears and the acquired image is