Quick Trouble Shooting - Kodak RVG 6500 IPS System User's Installation Manual

Wireless digital intra-oral x-ray system
Hide thumbs Also See for RVG 6500 IPS System:
Table of Contents

Advertisement

8
Troubleshooting
Occasionally, malfunctions can occur during use in the event of an incorrect action. The quick trouble
shooting and the Information "I", will guide you through the actions you need to take to correct the
malfunction.
Important: If the malfunction persists or more serious conditions
occur, contact your representative.
When you call your representative have the following information ready:
Model Number: KODAK RVG 6500
Error Code Number: I xx
Message displayed on the popup on the computer screen.

Quick Trouble Shooting

The quick trouble shooting guides you through actions you need to take to correct the malfunction.
The following table lists the malfunctions and the action to take.
Table 2
Quick Trouble Shooting
Malfunction
When you turn ON your RVG
sensor, the battery LED is red but
the control box is solid blue.
When you turn ON your RVG
sensor, it blinks blue then it turns
OFF.
When you turn ON your RVG
sensor, the battery LED is red and
the control box is blinking red.
Image quality is not good and
there are white lines on the image
When you turn ON your RVG
sensor, it blinks blue then after a
long time it becomes solid red.
Possible Cause
The battery needs
recharging.
The battery needs
recharging.
No battery
The battery needs
recharging.
WiFi access point is
not ON.
Configuration
problem.
KODAK RVG 6500 System_User & Installation Guide (SM728)_Ed01
Action
Recharge the battery.
You can continue acquiring images but the
quality will deteriorate.
Recharge the battery.
Recharge the battery
Recharge the battery
Check that the WiFi access point is
ON.
If it is ON and the problem persists
contact your representative
59