Figure 3-36: Bypass Installation; Table 3-18: Slg720 Bypass And Stillwell Recommended Diameters - Honeywell SmartLine SLG 700 User Manual

Level transmitter guided wave radar
Hide thumbs Also See for SmartLine SLG 700:
Table of Contents

Advertisement

Mounting on a bypass / bridle
SLG 700 transmitter can be successfully installed in a new or existing bypass pipe,
bridle, or a side pipe as shown Figure 3-36: Bypass installation. This type of installation
is often simpler and allows the addition of radar level measurement to an otherwise busy
installation.
A similar installation is also possible inside the main container, when installing the SLG
700 transmitter on a stilling well.
N = Inlet diameter
L = Effective measurement range (≥ 12"/300mm)
D = Bypass diameter (N<D)

Table 3-18: SLG720 bypass and stillwell recommended diameters

Probe type
Rod probe
Wire probe
Page 82

Figure 3-36: Bypass installation

Recommended diameter
3" or 4" (75mm or 100mm)
4" (100mm)
SLG 700 SmartLine Level Transmitter User's Manual
Minimum diameter
2" (50mm)
2" (50mm)
Revision 8