Figure 3-10: SLG726 flanged process connection, probe nut installation position, mm
Wire probe assembly
Wire probes can be supplied with an optional end weight.
Step
Fully thread the nut onto the central conductor. Using a lock washer, thread the
wire swage on to the central conductor. Torque the nut against probe and lock
washer to secure the connection.
Note:
1
Tighten the wire stud and nut to the following torque:
If applicable, insert wire probe into end weight. Tighten the 3 set screws to
secure end weight to wire probe
2
Note: Torque set screws to 6Nm (4.4ft-lbs)
Note: For flanged SLG726 models, ensure the nut does not intrude into the bore of the
process connector. See Figure 3-10: SLG726 flanged process connection, probe nut
installation position, mm [in] for more information.
Page 50
SLG720 6.0Nm (4.4ft-lbs)
SLG726 15Nm (11ft-lbs)