Surface Sgnl Str (R)
Surfac Sgnl Qlty (R)
Interfc Sgnl Str (R)
Interfc Sgnl Qlty (R)
Sensing Section (R)
Blk Dist Hi Zone (R)
Non-Crit
(single liquid only)
(continued)
Blk Dist Lo Zone (R)
(single liquid only)
Surface in BDH (R)
(two liquids only)
Surface in BDL (R)
(two liquids only)
Interface in BDH (R)
(two liquids only)
Interface in BDL (R)
(two liquids only)
Snsr Calibrated (R)
Calibration Type (R)
FldBkgdNotCompat
Backgrnd Not Set
FldgBkgdLoadError
Revision 8
Good
Bad
Good
Bad
Good
Bad
Good
Bad
OK
FAULT
Yes
No
Yes
No
Yes
No
Yes
No
Standard
Custom
Yes
No
Yes
No
Yes
No
SLG 700 SmartLine Level Transmitter User's Manual
Bad: Signal Strength is bad
Bad: Signal Quality is bad
Applicable for Two Liquid Products
Bad: Signal Strength is bad
Applicable for Two Liquid Products
Bad: Signal Quality is bad
This is a roll-up status that is set when any
of the following non-critical status
conditions are present:
 Sensor Over Temperature
 Level is in upper blocking distance zone.
 Level is in lower blocking distance zone.
 Sensor Not Characterized
Sensor Not Calibrated
These parameters mark regions outside
accurate measurement.
If the level enters either of these regions
then Status = Unknown.
These parameters mark regions outside
accurate measurement.
If the surface level enters either of these
regions then Status = Unknown.
These parameters mark regions outside
accurate measurement.
If the interface level enters either of these
regions then Status = Unknown.
Yes: Sensor is calibrated
No: Sensor is not calibrated
Standard: Device is calibrated for
standard points.
Custom: Device is calibrated for user-
defined points.
Yes: The Field Background is not
compatible.
Yes: The Background is not set.
Yes: Field Background loading error.