5.6 Trimming Coaxial Probes

Tools required

Required tools depend on options ordered.
For this item
Coaxial probe outer tube removal/installation
M3 set screw for coaxial coupler (SCA)
Inner rod probe
Inner rod probe nut
Mounting thread ¾" and 1" process connector
Mounting thread 1-½" process connector
Mounting thread 2" process connector
Coaxial probe outer tube threads (SCA)
Rod and tube trimming
Tool to drill hole in outer tube
1
Thread locking compound is recommended. The thread locking compound must be process compatible

Procedure

The Coaxial probe consists of an inner rod conductor and a coaxial outer tube shield. To trim the
coaxial probe to required length, both inner rod and coaxial outer tube need to be trimmed off by the
same amount.
Step – 1: Turn off transmitter power.
Step – 2: Remove transmitter from installed location.
Step – 3: Depending on the length of the coaxial probe the outer tube may be constructed from a
single length of tube or be assembled using several tube segments joined to each other using tube
couplers. The amount of probe to be trimmed will dictate the extent of disassembly required.
Starting from the end of the probe farthest from the process connector, remove the appropriate
number of outer tube sections to be able to complete the trimming operation. If the probe is under
2.0 m in length, it will be secured to the process connector by an internal thread. Use a wrench to
hold the process connector by its flats and a pipe wrench to unscrew the outer coaxial tube. If the
probe is over 2.0 m in length, the section of tube to be trimmed may be secured by a tube coupler.
To release the tube, loosen the two M3 set screws holding the coupler in place, then use two pipe
wrenches to unscrew the tube from the coupler.
Note: if only a small length of the probe requires trimming, only the end segment of the tube will
need to be removed from the assembly.
Revision 8
SLG 700 SmartLine Level Transmitter User's Manual
Use this tool
Pipe wrenches (2x)
AF 1.5 mm Allen key
AF 7 mm wrench (2x)
AF 8 mm wrench
AF 40 mm wrench
AF 50 mm wrench
AF 60 mm wrench
Process compatible thread locking
compound (i.e. Loctite 242)
Metal saw
Drill and 6.0 mm drill bit/drill press
1
Page 147

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents