Installation position

During the installation, please note that the instrument should be kept minimal distance to tank
wall from the vessel wall. For different antenna, please refer to
However, the instrument must not under any circumstances be mounted closer than 200mm to
the vessel wall or values calculated from
1. Reference plane
2. Center of the vessel or symmetry axis
Page 8
Figure 3-3: Installation position, >200 mm

Table 3-1: Minimal distance to tank wall

SLN700 Model
83A
82A 82B 83B 83C
82C 82D 83D 83E
87A 87B 87C 87D
SLN 700 SmartLine NCR Level Transmitter User's Manual
Table 3-1.
Min distance to tank wall
1/5 × Tank Height
1/10 × Tank Height
1/20 × Tank Height