Honeywell SmartLine SLG 700 User Manual page 134

Level transmitter guided wave radar
Hide thumbs Also See for SmartLine SLG 700:
Table of Contents

Advertisement

Device Vars
Display Info
(R) Read-Only parameter
<Return>
Firmware Version
Protocol (R)
Universal Rev (R)
(HART only)
Field Dev Rev (R)
(HART only)
Comm Info
Software Rev (R)
(HART only)
PV Alarm Type (R)
(HART only)
LRV (R) (FF only)
URV (R) (FF only)
<Return>
Firmware Version
Sensor Info
Model Key
Device ID
Page 118
Measurd Prod Lvl
Non-linearized
(R)
product level
Surface Sgnl Str
(R)
Signal Strength
Interfc Sgnl Str (R)
Signal Strength
For 2 liquid cases
Surface Sgnl Qlty
(R)
Signal Quality
Interfc Sgnl Qlty
(R)
Signal Quality
(two liquids only)
For 2 liquid cases
Int Elec Temp (R)
(two liquids only)
Temperature
Firmware Version
#
HART/FF
N/A
HIGH
LOW
N/A
N/A
N/A
SLG 700 SmartLine Level Transmitter User's Manual
<Return>
If linearization is disabled then Product
Level and Measurd Prod Lvl will be same.
Surface
Signal strength depends on a function of
distance and the medium. As the level and
interface peaks are negative, a more
Interface
negative value indicates a better signal.
Signal strength improves as distance
Normal range: (-500) – (-9000)
Surface
Quality measurements range:
Interface
Internal
Internal temperature measurement of the
Electronics
<Return>
#
Firmware version of display module
Firmware version of communications
module
Communications protocol of transmitter
Applicable only for HART device
Revision of Hart specification
Applicable only for HART device.
HART HCF hardware revision
For DD/DTM compatibility
Applicable only for Hart device
Hart HCF software revision
Applicable only for HART device
Loop current behavior during critical fault.
HIGH : 21.5mA
LOW : 3.5mA
Displays the analog output will be scaled
to 4mA.
Displays the measuring value for which
the analog output will be scaled to 20mA.
Firmware version of sensor module
Type and range of transmitter
Unique for each device considered as
serial number
HART only.
decreases.
0 – 1
Less than 0.3 = Bad
0.6 -- 0.8 = Good
1 = Perfect
sensing module.
Revision 8

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents