Installing the SINIUS CS dentist element
4.10.1
Installing the dentist element (standard position)
Installation
1. Place the bearing shells on the support column.
The pin (A) of the support column must be positioned in the groove
(B) of the bearing shell.
2. Fasten the bearing shells with an O-ring (C).
3. Insert the support tube in the support sleeve bearing-shells-first.
Important: The pin (A) of the support tube must be positioned in the
groove (D) of the support sleeve.
4. Align the support column using the 3 setscrews and a spirit level.
IMPORTANT
Required adjustments
The spring force of the flex arm must be adjusted following the complete
chair installation.
➢ Proceed according to section "SINIUS CS only: Adjusting the spring
force of the dentist element flex arm".