4
Startup
1.
2.
Reference Manual
Apply AC line power to the Rosemount
The probe takes approximately 45 minutes to warm up to the 1357 °F (736 °C)
setpoint. The 4-20 mA signal remains at a default value of 3.5 mA, and the O
reading remains at 0% through the warmup period.
After warmup, the probe begins reading oxygen, and the 4-20 mA output is based
on the default range of 0-10% O
If there is an error condition at startup, an alarm message is displayed on the
Rosemount CX1100 Remote Transmitter.
™
CX1100 Remote Transmitter.
.
2
Startup
2
21