The Rosemount CX1100 Remote Transmitter is available in a panel mounting or wall/pipe
mounting configuration. Refer to
mounting details. You need a wrench and bolts to mount the transmitter.
Procedure
1.
2.
3.
4.
Reference Manual
Ensure all components are available to install the Rosemount CX1100 Remote
Transmitter.
Select a mounting location near or removed from the Rosemount CX1100 Probe.
Consider the temperature limitations of the Rosemount CX1100 Remote
Transmitter. Refer to product specifications when selecting the mounting location.
Mount at a height convenient for viewing and operating the interface.
Emerson recommends approximately 5 ft. (1.5 m).
The keypad window on the Rosemount CX1100 Remote Transmitter may have an
exterior protective membrane. Remove the protective membrane prior to use of the
Rosemount CX1100 Remote transmitter.
Failure to remove the protective membrane may cause the display to appear
distorted. The membrane may become difficult or impossible to remove after
extended use at elevated temperatures.
Figure 2-4
or
Figure 2-5
for the panel, wall, or pipe
Install
11