Download Print this page
Emerson Rosemount CX1100 Quick Start Manual
Emerson Rosemount CX1100 Quick Start Manual

Emerson Rosemount CX1100 Quick Start Manual

In situ oxygen transmitter
Hide thumbs Also See for Rosemount CX1100:

Advertisement

Quick Links

Quick Start Guide
00825-0100-4110, Rev AB
April 2020
Rosemount
CX1100
In Situ Oxygen Transmitter

Advertisement

loading

Summary of Contents for Emerson Rosemount CX1100

  • Page 1 Quick Start Guide 00825-0100-4110, Rev AB April 2020 ™ Rosemount CX1100 In Situ Oxygen Transmitter...
  • Page 2 • Read all instructions prior to installing, operating, and servicing the product. • If you do not understand any of the instructions, contact your Emerson representative for clarification. • Follow all warnings, cautions, and instructions marked on and supplied with the product.
  • Page 3 When replacement parts are required, ensure that qualified people use replacement parts specified by Emerson. Unauthorized parts and procedures can affect the product's performance, place the safe operation of your product at risk, and void your warranty. Look-alike substitutions may result in fire, electrical hazards, or improper operation.
  • Page 4: Table Of Contents

    Restrict physical access by unauthorized personnel to protect end users’ assets. This is true for all systems used within the facility. Contents Install............................5 Wire............................17 Start up............................22 Calibration..........................23 Product certifications......................... 25 Declaration of Conformity......................27 China RoHS table........................28 Emerson.com/Rosemount...
  • Page 5: Install

    Do not allow the temperature of the probe electronics to exceed 194 °F (90 °C) Install probe Emerson can supply a weld plate for welding to the flue gas duct for new installations. Procedure 1. If using the standard square weld plate...
  • Page 6 D. ½ national pipe thread (NPT) conduit connection (power, signal) E. Breather port F. Calibration gas connection Table 1-1: Removal/Installation Probe length Removal envelope .5 m (19.68 in.) 750 mm (30 in.) 1 m (39.37 in.) 1250 mm (49 in.) 2 m (78.74 in.) 2250 mm (89 in.) Emerson.com/Rosemount...
  • Page 7 April 2020 Quick Start Guide Figure 1-2: Probe Installation All dimensions are in millimeters with inches in parentheses. Square weld plate, ANSI pattern part 4512C34G01 Quick Start Guide...
  • Page 8 4 holes equally spaced on 120.7 mm (4.75 in.) 145 mm (5.71 in.) Figure 1-3: Drip Loop and Insulation Removal Note Replace insulation after installing Rosemount CX1100. B. Drip loop C. Stack duct or metal wall D. Adapter plate E. Insulation Note Probe installation may be vertical or horizontal.
  • Page 9 April 2020 Quick Start Guide Mount remote display The remote transmitter is available in a panel mounting or wall/pipe mounting configuration. Prerequisites You need a wrench and bolts to mount the transmitter. Note Dimensions are in inches with millimeters in parentheses. Figure 1-4: Wall/Surface Mount Front View A.
  • Page 10 Quick Start Guide April 2020 Figure 1-5: Wall Mount Side View Emerson.com/Rosemount...
  • Page 11 April 2020 Quick Start Guide Figure 1-6: Pipe Mount Bottom View A. Front panel B. Six ½-in. (12.7 mm) national pipe thread (NPT) conduit openings C. Mounting bracket D. U-bolts Quick Start Guide...
  • Page 12 Quick Start Guide April 2020 Figure 1-7: Pipe Mount Side View A. 2-in. (50.8 mm) pipe supplied by customer Emerson.com/Rosemount...
  • Page 13 April 2020 Quick Start Guide Figure 1-8: Panel Mount Front View The front panel is hinged at the bottom. The panel swings down for easy access to the wiring locations. Quick Start Guide...
  • Page 14 Quick Start Guide April 2020 Figure 1-9: Panel Mount Side View A. Maximum panel thickness 0.375 in. (9.52 mm) B. Panel mount gasket C. Four mounting brackets and screws provided Emerson.com/Rosemount...
  • Page 15 April 2020 Quick Start Guide Figure 1-10: Panel Mount Bottom View A. Six ½-in. (12.7 mm) NPT conduit openings Figure 1-11: Panel Cut-Out Procedure 1. Ensure all components are available to install the remote transmitter. Quick Start Guide...
  • Page 16 3. Mount at a height convenient for viewing and operating the interface. Emerson recommends approximately 5 ft. (1.5 m). 4. The keypad window on the transmitter may have an exterior protective membrane. Remove the protective membrane prior using the transmitter.
  • Page 17: Wire

    April 2020 Quick Start Guide Wire All wiring must conform to local and national codes. WARNING Failure to follow safety instructions could result in serious injury or death. Before installing the equipment, read the safety instructions at the front of this document.
  • Page 18 2. Remove the covers from the probe and the remote transmitter (if applicable). 3. Feed all probe wiring through the conduit port of the probe. 4. Connect probe heater power leads to probe connector. Refer to Figure 2-1. Emerson.com/Rosemount...
  • Page 19 April 2020 Quick Start Guide Figure 2-1: Closeup Wiring Diagram A. Labels B. M4 X 0.7 X 8 pan head screw machine screw (internal ground) 5. Connect O signal, thermocouple, and cold junction (CJC) wires from the probe to the transmitter 6.
  • Page 20 Figure 2-2: Wiring Head A. 4-20 mA out B. Heater power to probe C. Power supply D. Combustion sensor E. Shield ground F. Alternate 4-20 mA out G. Signal from probe H. Power to remote transmitter I. Alarm relay Emerson.com/Rosemount...
  • Page 21 April 2020 Quick Start Guide Figure 2-3: Wiring Boards A. AC power wires from power supply board to the underside of the sensor board are provided with the sensor board. B. AC input C. Heater power: terminates to underside of the sensor board. D.
  • Page 22: Start Up

    2. After warm-up, the probe begins reading oxygen, and the 4-20 mA output is based on the default range of 0 to 10 percent O If there is an error condition at startup, an alarm message is displayed on the remote transmitter. Emerson.com/Rosemount...
  • Page 23: Calibration

    April 2020 Quick Start Guide Calibration You can calibrate the Rosemount CX1100 in the installed condition without removing the instrument from the process duct and also while the combustion process is online. Gas is applied to the sensor through the calibration gas fitting. New Rosemount CX1100 systems are factory calibrated and are generally acceptable for initial start-up and operation.
  • Page 24 April 2020 9. Follow the instructions on the display, manually applying calibration gases to the calibration port. Emerson recommends applying 0.4 percent and 8.0 percent O at 5 SCFH. 10. After the calibration is finished, press Left and use the Up and Down buttons the check the current and previous ten calibration logs.
  • Page 25: Product Certifications

    A copy of the EU Declaration of Conformity can be found at the end of the Quick Start Guide. The most recent revision of the EU Declaration of Conformity can be found at Emerson.com/Rosemount. Ordinary location certification As standard, the transmitter has been examined and tested to determine...
  • Page 26 4. The probe unit of the equipment shall be connected to protective earth separately at end installation. 5. End installation to provide the means of disconnection from power sources. 6. End installation to comply with the requirement of IP66 and 4X. Emerson.com/Rosemount...
  • Page 27: Declaration Of Conformity

    April 2020 Quick Start Guide Declaration of Conformity Quick Start Guide...
  • Page 28: China Rohs Table

    Quick Start Guide April 2020 China RoHS table Emerson.com/Rosemount...
  • Page 29 April 2020 Quick Start Guide Quick Start Guide...
  • Page 30 Quick Start Guide April 2020 Emerson.com/Rosemount...
  • Page 31 April 2020 Quick Start Guide Quick Start Guide...
  • Page 32 Linkedin.com/company/Emerson- Automation-Solutions The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a twitter.com/rosemount_news mark of one of the Emerson family of companies. Facebook.com/Rosemount All other marks are the property of their youtube.com/RosemountMeasurement respective owners.