Optimizing Measurements; Avoid Stray Light; Shielding The Sensor; Filtering And Attenuation - Newport CONEX-PSD User Manual

Two-axis position & power sensing device
Table of Contents

Advertisement

CONEX-PSD
4.0

Optimizing Measurements

This section provides general recommendations for improving light measurement
setups.
4.1

Avoid Stray Light

As the sensitive area of a PSD is relatively large (81 mm
may collect unwanted light on top of the beam being measured. This may affect the
measurement accuracy.
PSD's are also sensitive to a large spectral range (300 to 1100 nm for Si and 800 to
1700 for Ge). Although the light to be measured might be centered on one particular
wavelength, sensor will be affected by other wavelengths.
For example, stray light can easily reach 10 μW/mm
room. This might not be negligible, especially if a large gain was set in the controller.
There are 2 ways to minimize the effect of stray light:
• Shield the sensor from unwanted light.
• Block the unwanted wavelength.
4.1.1

Shielding the Sensor

To minimize this effect, it is recommended to shield the sensor from stray light.
One or several Newport filter holders can be used for that.
4.1.2

Filtering and Attenuation

Several types Newport filter can be mounted on the Newport "Model 883-OH 1" filter
holder.
For example:
Absorptive Neutral Density Filter (FSR-xxxx):
Two-Axis Position & Power Sensing Device
9
2
for Si, 100mm² for Ge), it
2
on a broad spectral range in a
EDH0315En1030 — 07/16

Advertisement

Table of Contents
loading

Table of Contents