Analog and power board

Connecto
Function
r
J1
SIG 1. Analog output for signal 1.
J3
REMOTE. Synchronize start/stop functions
for up to 10 devices.
J5
96-pin data connection to analog and
power board.
J7
Receive power from transformer
secondary.
J8
Ribbon cable to automatic liquid sampler
board.
P1
Supply power to the Communication
harness.
P18
Connects to the Motor Harness and AC
Control cable.
P21
Connects to Inlet/Detector harness.
P22
Connects to the Auxiliary Zone/Valve Box
Harness.
Agilent 7820 GC Service
287
Part number G4350-61150.
See also
Analog signal outputs (
178)
Remote Start/Stop ( 172)
Main transformer ( 394)
RS-232 cable, ALS
controller
Communication harness
Motor Harness & AC Control
Assembly
Inlet/Detector harness
Auxiliary zone/Valve box
harness