Replacing The Fpd Signal Board - Agilent Technologies 7820A Service Manual

Gas chromatograph
Hide thumbs Also See for 7820A:
Table of Contents

Advertisement

Agilent 7820 GC Service
123
Wait 20 minutes for the detector to heat up, then ignite the flame.
Long cable connectors and leads
Detector type
Single-waveleng
th
Single-waveleng
th

Replacing the FPD signal board

If changing the GC configuration, see Changing the GC
configuration ( 167) for important information regarding GC
methods. Then proceed with the steps below.
1 Cool down the oven and all heated zones.
Turn off all gas flows at their sources.
Turn off the GC main power switch and disconnect its power
cord.
Location Connecto
r
Front
A1
Back
A1