Power diagnostics
Agilent 7820 GC Service
281
This information is based on Service Mode | Diagnostics |
Electronics | Power Info.
POWER DIAGNOSTICS
Front inl:
72.5W
Front det:
74.2W
Aux 1:
0.5W
Back inl:
72.5W
Back det:
74.2W
Logic current
Valve current
24 Volts
-24 Volts
Zone supply
Primary
voltage
The examples were taken from a GC with no auxiliary heaters
configured.
Note that the power display shows real-time percentages.
39.8
****%
C
40.8
****%
C
----
****%
----
C
39.8
****%
C
40.8
****%
C
1.37x
amps
0.176
amps
26.66
volts
-27.03
volts
42.92
volts
99.7%