To Check FPD Performance (Sample 5188-5953)

150
Agilent 7820 GC Service
If sample FPD checkout sample 5188-5953 is available, use it to
test FPD performance.
To check FPD performance, first check the phosphorus
performance, then the sulfur performance.
If you are using OpenLab ChemStation edition or GC
ChemStation software with your GC, the testing process can be
automated using two macros. The macros are available on the
CE-ROM and from the EPI Warehouse and the Agilent Website.
Macro name
MEPCK_S.MAC
MEPCK_P.MAC
CE-ROM
Look in Gas Chromatography / Patches and Firmware / FPD
Checkout Macros.
EPI Warehouse
The download page is:
http://whadmin.cos.agilent.com/search/QuickResults.asp?s
Pub=True
.
Agilent Website
The download page is:
http://www.chem.agilent.com/en-US/Technical-Support/Soft
ware-Informatics/ChemStation-for-GC-LC-CE-A-D-LC-MS/Util
ities/Pages/fpdcheckout51885953.aspx
Copy the macros into the Core directory of the ChemStation.
This is usually either C:\CHEM32\CORE or C:\HPCHEM.
As a post-run command
Enter the command on the Run-Time Checklist screen. The
Description
Processes data collected from sulfur
tests.
Processes data collected from
phosphorus tests.