To Check uECD Performance

Agilent 7820 GC Service
145
output, Range 0)
10
7
1
1000
C , C , C
area counts 525000
C /C
area ratio
1 Gather the following:
Evaluation column, HP-5 30 m — 0.32 mm — 0.25 µm
(19091J-413)
uECD performance evaluation (checkout) sample
(18713-60040, Japan: 5183-0379)
4-mL solvent and waste bottles or equivalent for
autoinjector
Chromatographic-grade isooctane
2-mL sample vials or equivalent for sample
Verify the following:
Clean fused silica indented mixing liner installed. If not,
install it.
Chromatographic-grade gases plumbed and configured:
helium as carrier, nitrogen as makeup. (Nitrogen is an
acceptable alternate carrier gas.)
Empty waste vials loaded in sample turret.
4-mL vial with diffusion cap filled with hexane and inserted
in Solvent A injector position.
Install the evaluation column. (See the procedure for the SS or
PP in the Maintenance manual.)
Bake out the evaluation column for at least 30 minutes at
Pk wd
Thresh
Stop time
SS
1.00 +/-
0.10
0.04
5
8.75
PP
460000
1.00 +/-
0.10