Configuration locks

168
Agilent 7820 GC Service
change in configuration and force you to resolve the differences.
If you do not resolve differences in both the configuration and
method parameter screens the method will not load. Run method
or run sequence will not start the run.
Use the GC Configuration Connection screen to get the new
configuration. Review and update any configuration setpoints.
Use the Edit parameters screen to review and update any
method setpoints.
Methods stored in GC
Methods stored in the GC that do not match the new
configuration must be updated. Make the necessary additions or
modifications. Store this method over the old method.
When you load a method, the GC compares the method against
the current hardware it detects (inlet types, detector types, PCM,
heaters, valves, and so forth). A method which does not match is
not loaded. Default setpoints for this configuration are loaded.
Before adding new hardware or changing the installed hardware
type, verify that any methods stored in the GC have been
documented.
When trying to change the GC configuration—as by adding,
removing, or changing an Inlet module—you may encounter this
or a similar message:
Configuration is Locked
Go to keyboard options to unlock
To unlock the configuration
1 Press [ Options ].
Select Keyboard and Display .
Select Hard Configuration lock , then press [ No/Off ].
When you power cycle the GC, the configuration lock is
automatically set to On.
When you use the Reboot GC feature under Options |
Communications, the configuration lock is not changed.