How to replace the old version uECD for customer
7820 old version uECD is not an orderable part. If customers
are experiencing failures on 7820 old version uECD and the
uECD cell must be replaced, the new version uECD (P/N
G2397-60620) and two unions (P/N 0100-0124) can be
ordered for uECD cell replacement. Refer to below for parts
required and replacement procedure.
Description
7820A replacement ECD cell kit, no
signal board, no EPC module, no ship kit
(New version detector)
Union SS 1/16inch tubing
1 Cut anode purge and make-up gas tubes between thumb
nuts and old version uECD body. See below figure.
2 Uninstall old version uECD body from GC.
3 Cut two tubing blocks for anode purge and makeup from the
new version uECD, P/N G2397-60620.
4 Install new version uECD body without tubing blocks on GC
Part number
G2397-60620
0100-0124