N5280A
General Specifications
Dimensions and Space Requirements
Standard installation of the N5280A and PNA includes configuration and
installation on a customer provided lab bench, or table top of adequate size
and strength.
Table 5
Model
N5242A
N5264A
N5280A
Maximum Power Levels
Refer to your PNA specifications to determine the maximum input power
levels for the PNA access and test ports, or to optimize the power levels in
the receivers.
Damage and maximum levels are not necessarily the optimum level.
14
Instrument Dimensions
Weight
Height
37 kg
26.67 cm
82 lb (±0.5 lb)
(10.5 in)
22 kg
26.67 cm
48 lb (±0.5 lb)
(10.5 in)
9.1 kg
8.9 cm
(20 lb)
(3.5 in)
Table 6
Power Levels
RF Input Power Damage Levels
RF Port
+18 dBm
LO Port
+5 dBm
Optimum LO Power
0 dBm (±1 dB)
RF Input @ 0.1 dB Typical Compression
10 dBm
A- D Receiver
1
IF Output Level
Ports A- D
10 dBm
1.
IF Output level is based on the RF Input @
0.1 dB typical compression.
Wid th
Depth
55.8 cm
42.5 cm
(16.75 in)
(21.97 in)
55.8 cm
42.5 cm
(16.75 in)
(21.97 in)
42.5 cm
48.3 cm
(16.75 in)
(19 in)
Keysight N5280A User's and Service Guide