Operational Check
Table 15
Measured IF OUT Power Levels
PSG #1
RF IN
Output Freq
Power
to RF IN A
Level (dBm)
(GHz)
10 dBm
8.5
17.5
26.0
Keysight N5280A User's and Service Guide
PSG #2 
LO IN 
Output Freq
Power
to RF IN A
Level (dBm)
(GHz)
0 dBm
9
18
26.5
Power Level
Power Meter Read ing Measured
Range Expected
for IF OUtputs and LO AUX (dBm)
at IF OUT A
(dBm)
A
B
5 to 11
5 to 11
5 to 11
C
D
LO AUX
41