Product description
Setting the logs via software
To set the logs using software please refer to the HMS CS operating manual for the respective
piece of patient data management software.
Note:
▪
Logs 1, 2.10 and 11 can be changed via the HMS CS evaluation software.
The changeable parameters are:
1.
Display of readings on the display
2.
Start of the 4 day intervals and the number of measurements per hour. The following are
available: 4,5,6,10,12,15,20,30 per hour
3.
Turning on and off function of the buzzer for the 4 day intervals
▪
Logs 3 to 9 are pre-set and cannot be modified.
▪
Log 5 is suitable for activities during the night (night shift).
▪
Log 9 is the Schellong test. Here, the blood pressure is measured every 2 minutes during the first
30 minutes. After 30 minutes, the Mobil-O-Graph
▪
Log 10 enables the automatic transmission of readings via Bluetooth
actual measurements. To activate the Bluetooth
to blood pressure measurement, log 10 allows you to integrate pulse wave analysis with the Mobil-
O-Graph
®
PWA.
▪
Log 11 is used for 24-hour pulse wave analysis, and can only be implemented with the Mobil-O-
®
Graph
PWA. The measuring intervals are set via the HMS CS.
90
®
takes measurements with log setting 3.
®
directly to the doctor's PC for
®
button you need a Bluetooth interface. In addition
Need help?
Do you have a question about the Mobil-O-Graph PWA and is the answer not in the manual?
Questions and answers
Que veux dire code 6 et code 2
Code 6 means that there are no unsent blood pressure values in the measurement memory of the Mobil-O-Graph®.
Code 2 indicates that the Bluetooth interface of the Mobil-O-Graph® could not be configured properly, possibly due to a communication problem between the device and the Bluetooth module.