IEM Mobil-O-Graph PWA Operating Manual page 76

Long-term blood pressure measuring device
Hide thumbs Also See for Mobil-O-Graph PWA:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 68
Directions for use
WARNING
Risk of strangulation by shoulder strap and cuff tube
If the patient has limited cognitive abilities, the device may only be used under supervision.
The Mobil-O-Graph
If the Mobil-O-Graph
constant supervision of the doctor in accordance with regulations.
Do not place the shoulder strap and cuff tube around the patient's neck.
Instruct the patient to wear the cuff only on the upper arm and, in any case, to make sure that neither
the shoulder strap nor the pressure tube could ever wrap around the neck. For this purpose, the air
tube is always laid under the outer clothing (even at night).
Be sure to tell patients about the correct positioning of the cuff.
Instruct the patient to turn off the device, remove the cuff and inform you if they are experiencing any
pain, swelling, redness or numbness in the arm around which the cuff is placed. (It is expected that
the patient may experience some mild to moderate discomfort during a blood pressure measurement.)
Measurement can be interrupted at any stage by pressing any button. This deflates the cuff and the
device can be removed.
WARNING
Risk of injury from putting on and pumping up a cuff over a wound
Do not place the Mobil-O-Graph
74
®
must not be used on patients of unsound mind.
®
is used on children, this should be done with particular caution and under the
®
over a wound or bandage.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Mobil-O-Graph PWA and is the answer not in the manual?

Questions and answers

Biunkens sonja
February 6, 2025

Que veux dire code 6 et code 2

1 comments:
Mr. Anderson
February 10, 2025

Code 6 means that there are no unsent blood pressure values in the measurement memory of the Mobil-O-Graph®.
Code 2 indicates that the Bluetooth interface of the Mobil-O-Graph® could not be configured properly, possibly due to a communication problem between the device and the Bluetooth module.

This answer is automatically generated

John colmenares
February 27, 2025

Calibración

User image 67c0a4f19a301