IEM Mobil-O-Graph PWA Operating Manual page 122

Long-term blood pressure measuring device
Hide thumbs Also See for Mobil-O-Graph PWA:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 68
Appendix
WARNING
After an automatic measurement, allow at least 3 minutes to elapse to avoid prolonged
impairment of blood circulation before actively starting a measurement.
Audible signals
The audible signals generated by the device consist of individual or multiple tone burst sequences. The following
tone burst sequences are emitted:
Tone burst
1 tone burst
3 tone bursts
Continuous tone bursts
120
Heard when
Switching on and off
Starting and ending measurement (except during the night-time
interval)
Removing the interface cable, ending IR communications,
establishing and ending Bluetooth communications
Measurement errors
System error
Serious system error (e.g. the cuff pressure is greater than 15 mmHg
for more than 10 seconds outside the measurement)

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Mobil-O-Graph PWA and is the answer not in the manual?

Questions and answers

Biunkens sonja
February 6, 2025

Que veux dire code 6 et code 2

1 comments:
Mr. Anderson
February 10, 2025

Code 6 means that there are no unsent blood pressure values in the measurement memory of the Mobil-O-Graph®.
Code 2 indicates that the Bluetooth interface of the Mobil-O-Graph® could not be configured properly, possibly due to a communication problem between the device and the Bluetooth module.

This answer is automatically generated

John colmenares
February 27, 2025

Calibración