Troubleshooting
Error message
Err 3
Err 4
104
Possible cause
1. Blood pressure is beyond the
measurement range
2. Considerable arm movement
3. Problems with the pneumatics
1. Data transmission cable not
correctly inserted into the blood
pressure monitor
2. Pins in the plug of the data
transmission cable are
mechanically damaged.
3. Reading not transmitted
correctly
Measures
1. Permanent notifications render
the blood pressure monitor
unsuitable for the patient.
2. Keep the arm still during
measurement.
3. If this error occurs continuously,
send the device to your specialist
supplier or directly to IEM GmbH
for checking.
1. Insert the cable correctly into the
blood pressure monitor (see
chapter 3.4 "Preparation for
measurement").
2. Check the plug to see whether
the pins on the inside are
damaged. If they are, call your
specialist supplier or IEM
3. Start the transmission again.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Mobil-O-Graph PWA and is the answer not in the manual?

Questions and answers

Biunkens sonja
February 6, 2025

Que veux dire code 6 et code 2

1 comments:
Mr. Anderson
February 10, 2025

Code 6 means that there are no unsent blood pressure values in the measurement memory of the Mobil-O-Graph®.
Code 2 indicates that the Bluetooth interface of the Mobil-O-Graph® could not be configured properly, possibly due to a communication problem between the device and the Bluetooth module.

This answer is automatically generated