4.2.14 OUTPUT CAL FACTORS Menu
The OUTPUT CAL FACTORS menu provides access to the output calibration parameters.
These parameters are password protected and can only be changed after the calibration
password has been entered. For three phase configurations, the PHASE keys toggle between
the three calibration screens for each phase. Refer to the calibration section in this manual for
details on performing a calibration.
The following calibration factors are available from this menu:
Entry
VOLT FS
V HI-FREQ
VOLT ZERO
PHASE OFST
IMP. REAL FS
IMP. REACT FS
IMP. REAL MIN
IMP. REACT MIN
76

Figure 4-27: OUTPUT CAL FACTORS menu

Description
Full scale voltage output calibration factor.
High frequency full scale voltage output calibration factor.
Zero offset voltage calibration factor.
Phase offset calibration factor. Compensates for phase shift
caused by AC amplifier.
Full scale resistive output impedance calibration factor.
Full scale inductive output impedance calibration factor.
Minimum resistive AC source output impedance. The AC source
has an output impedance greater than zero. This value
determines the minimum resistive component of the AC source
output impedance.
Minimum reactive AC source output impedance. The AC source
has an output impedance greater than zero. This value
determines the minimum inductive component of the AC source
output impedance.
i Series / iX Series