4.6.2.2 Analyzing waveform data
The data displays available for acquired waveform data allow you to scroll through the entire
acquisition buffer. For waveform displays, the knob can be used to scroll through the display
horizontally. The UP and DOWN cursor keys have no effect in this display mode.
Readouts on the
left track the
vertical cursor
position on the
waveform.
Trigger point is
at 0.0 ms.
The acquisition buffer may be longer than the selected horizontal scale in which case only a
portion of the acquisition window will be visible. The left portion of the LCD display is used to read
out the data under the vertical cursor. This cursor is a dotted line that can be moved using the
knob. If the left or right edge of the display window is reached, the entire display will shift so the
cursor always remains visible. The time from the trigger point to the vertical cursor is displayed in
the left hand portion of the LCD in ms. Also displayed here are the absolute voltage and/or
current value under the cursor.
To change the horizontal display scale without re-triggering an acquisition, press the ENTER key
to return to the HARMONICS/TRACE ANALYSIS screen, change the SCALE value, move the
cursor to the VIEW field and press ENTER. This will display the same data set at the new scale
without triggering a new acquisition.
88

Figure 4-35: Scrolling through acquired waveform data

Counter
Clock
Clock
wise
wise
DECR
INCR
i Series / iX Series