150 VDC Range Volt Zero:
Press the PROGRAM key. Select VOLT RANGE and program the 150 VDC range with the
shuttle. Go to the OUTPUT CAL screen and adjust the VOLT ZERO for 0.0 0.005 volts DC on
the output.
150 VDC Range + Full-scale:
Program +120.0 volts. Go to the OUTPUT CALIBRATION screen and adjust the VOLT FS
parameter for an output voltage of +120.0 ±0.05 volts DC.
150 VDC Range - Full-scale:
Program -120.0 volts. Go to the OUTPUT CALIBRATION screen and adjust the VOLT FS
parameter for an output voltage of -120.0 ±0.05 volts DC.
For a 9003iX or 15003iX 3-Phase power system with one controller, repeat the preceding steps
for the Phase B and C outputs. Press the PHASE key to select the output to be calibrated.
Monitor the output of the respective phase by moving the HI input of the Digital Multimeter. The
LO input should remain connected to the common LO of the sense connector, TB3.
The following Table is a summary of the preceding calibration steps.
Program the following values in the table and make the adjustments in the OUTPUT
CALIBRATION screen. Select the phase to be calibrated by pressing the PHASE key.
TITLE
150 VAC range DC Zero
150 VAC range Volt FS
150 VAC range Volt Hi Freq
Repeat the adjustments at 60 and 400 Hz until the output is within ± 0.05 volts
300 VAC range DC Zero
300 VAC range Volt FS
300 VAC range Volt Hi Freq
Repeat the adjustments at 60 and 400 Hz until the output is within
150 VDC range DC Zero
150 VDC range Volt +FS
150 VDC range Volt -FS
300 VDC range DC Zero
300 VDC range Volt
300 VDC range Volt -FS
For a multi-phase power system that uses one controller, 9003iX or 15003iX, repeat paragraph
6.4 for each phase. Move the external multimeter to the phase being calibrated. Make sure that
the phase being calibrated is programmed to the correct voltage. While viewing the calibration
screen, press the PHASE key to select the respective key.
116
PROGRAM VALUES
CALIBRATION VALUE
150 range, 0.0 V
VOLT ZERO
120.0 V, 60 Hz
VOLT FS
120.0 V, 400 Hz
V HI FREQ
270 range, 0.0 V
VOLT ZERO
240.0 V, 60 Hz
VOLT FS
240.0 V, 400 Hz
VOLT HI FREQ
135 range, 0.0 V
VOLT ZERO
+120.0 V
VOLT FS
-120.0 V
VOLT FS
270 range 0.0 V
VOLT ZERO
+240.0 V
VOLT FS
-240.0 V
VOLT FS

Table 6-1: Output Calibration Table

ADJUST TO
0 ± 5 mv DC
120 0.05 VAC
120
0.05 VAC
0 ± 5 mv DC
240
0.05 VAC
240
0.05 VAC
0.05 volts
0 ± 5 mv DC
+120
0.05 VDC
-120
0.05 VDC
0 ± 5 mv DC
+240
0.05 VDC
-240
0.05 VDC
i Series / iX Series