Autoclaving

The Transferpette
and 20 min. exposure time ( t
1.
Eject the pipette tip.
2.
Autoclave the complete pipette without
any further disassembling.
3.
Allow the Transferpette
cool and dry.
Note:
The effectiveness of the autoclaving must
be verified by the user.
Maximum reliability is obtained with vacuum
sterilization. We recommend the use of
sterilization bags.
Attention:
During autoclaving, the volume setting
lock must be in the upper switch position
(UNLOCK)!
If the pipette is autoclaved frequently, the piston
should be oiled with the supplied silicone oil in
order to preserve smooth movement.
If necessary after sterilization, tighten the con-
nection between the hand grip and the pipette
shaft.
5 ml + 10 ml Filter Transferpette
A hydrophobic PE filter is used as a safeguard
against liquid entering the pipette.
Change the filter if it becomes wet or contaminated.
• Use a flat object such as a screwdriver.
• Remove the filter without damaging the tip cone.
Remove the filter before autoclaving!
The instrument can be operated without a filter.
30
®
S is completely autoclavable at 2 bar (121°C)
) according to DIN.