Keysight 16197A Bottom Electrode SMD Test Fixture
Operation and Service Manual
4

User Maintenance

This chapter describes the operations to perform before starting measurement,
and the procedures necessary to ensure accurate measurements.

Description

Necessity of User Maintenance

Through repeated use, the fixture's measuring performance will gradually
deteriorate over time due to smearing on the contact faces from solder
buildup, and to mechanical wear and deformation of the contact faces
themselves. To continually achieve optimal measurement results, it is
important to keep the contact faces in good condition and take appropriate
action before such wear or deformation occurs. Therefore, the fixture should be
properly cared for by implementing each of the maintenance items shown in
Table
Table 4-1
Maintenance Items
Name
Cleaning
Electrode
deterioration
check
4-1.
Frequency of
implementation
Several times a day
Before using the
fixture for the first time
after delivery/ After
replacement of parts
Once a day/Before
fixture compensation
Contents
Cleaning the
fixture
Acquisition of
reference
a
values
Determining the
deviation from
reference values
43
Target parts
Entire fixture
Center electrode,
Electrode plate
Center electrode,
Electrode plate
a

Advertisement