Operation
Performing Fixture Compensation
Figure 3-17
Standby Position of the Pressure Road
Confirm that the end of the pressure rod is not deformed. Replace the rod
if deformed. For the replacement part, see
page
Measuring OPEN Compensation Data
With the fixture in an OPEN state, measure the OPEN compensation data.
Step 7. Press the lever to raise the pressure rod, and remove the shorting bar used to
measure the SHORT compensation data
Step 8. Press the lever to raise the pressure rod
secure the pressure rod in its raised position
Figure 3-18
Setting the Fixture for OPEN State
38
61.
"Replaceable Parts" on
(Figure
3-18,1).
(Figure
3-18,2), and push the latch to
(Figure
3-18,3).