Check Sheet; Example Entry In Check Sheet - Keysight Technologies 16197A Operation And Service Manual

Bottom electrode smd test fixture
Table of Contents

Advertisement

User Maintenance

Check Sheet

Check Sheet

Example Entry in Check Sheet

The following tables shown an example entry of electrode deterioration check
results in the Check Sheet.
Example Entry of Electrode Deterioration Check Results
DUT Size
2012
Table 4-2
Example Entry of Reference Values and User Limit Values
Frequency
100 MHz
800 MHz
Table 4-3
Example Entry of Checking History
Date/
Time
2000/
10/11
9:30
2000/
10/11
9:35
2000/
10/12
9:30
2000/
10/12
9:35
Keysight16197A Operation and Service Manual
Measurement
a
parameter
Rs
Ls
Rs
Ls
a. User's discretionary value
b. Enter the values obtained in section
page 48
"Example Settings of User Limit Values" on page 47
c. For entry, see
Frequency
Measurement
parameter
100 MHz
Rs
Ls
800 MHz
Rs
Ls
100 MHz
Rs
Ls
800 MHz
Rs
Ls
a.
Replace parts if check results are not acceptable.
Reference
User limit
b
c
value [a]
value [b]
40 m
30 m
2.34 nH
0.5 nH
220 m
40 m
2.32 nH
0.4 nH
"Acquiring Reference Values" on
Measurement
value
50 mW
2.5 nH
250 mW
2.6 nH
55 mW
2.6 nH
285 mW
2.7 nH
Lower limit
Upper limit
[a-b]
[a+b]
10 m
70 m
1.84 nH
2.84 nH
180 m
260 m
1.92 nH
2.72 nH
Acceptability
Set position
of pressure
adjusting
nut
O
O
O
O
O
O
a
X
O
51