Changing the Orientation of the Device Guide and Electrode Plate

Changing the Orientation of or Replacing the Device Guide and Electrode Plate

Change the orientation of the device guide and electrode plate so that the
selected device guide frame is located at the measuring position. For a 0603
device, replace the attached device guide and electrode plate with ones
supplied in Option 001.
Step 1. Loosen the knobs on the sliding block, and remove the pressure unit from the
electrode unit, referring to
Step 2. Unscrew the screws that fix the guide holder. Then remove the guide holder,
device guide, and electrode plate
Figure 3-4
Removing the Device Guide and Electrode Plate
If it is difficult to take out the electrode plate, turn the electrode unit
upside down to remove the plate. Do not use force to take out the
electrode plate, otherwise it may be deformed.
Step 3. Mount the device guide and electrode plate so that the selected device guide
frame is located at the measuring position. Ensure that the marks on the
device guide are aligned with those on the electrode plate
0603 DUT, replace the attached device guide and electrode plate with ones
supplied in Option 001
28
Figure 1-4 on page 9
(Figure
3-4)
(Figure
3-6).
Keysight16197A Operation and Service Manual
(Figure
3-5). For a