7 Maintenance

7.2 Quality Control

The Quality Control (QC) programs in the Quality Control menu are used to ensure that
the image quality remains constant.
Quality Control should be performed at regular intervals, preferably at least once a
month and always after calibration.
NOTICE! The device will automatically remind of retaking the QC images. If
you want the change the frequency of the reminders, contact your service for
assistance.
NOTICE! QC programs produce X-rays. Protect yourself from radiation.

7.2.1 PAN QC

NOTICE! The PAN QC is an optional, yet recommendable, procedure and is
mandatory to be performed where local regulations require it.
1. GUI: Go to Settings menu.
2. GUI: Select Quality Control menu.
3. GUI: Select PAN QC program from the list.
4. Attach the Panoramic QC phantom holder and 2D QC phantom to the device.
5. Attach the copper filter in front of the radiation window on the tube head. The filter
attaches in place with magnets.
44