Geometry Calibrations - KaVo ORTHOPANTOMOGRAPH OP 3D User And Installation Manual

Orthopantomograph
Hide thumbs Also See for ORTHOPANTOMOGRAPH OP 3D:
Table of Contents

Advertisement

9 Installation
3. Press and hold down the exposure button to take the calibration image.
4. When the exposure warning stops and the program end tone is played, the program
is complete.
5. The calibration image appears on the GUI.
6. Acknowledge the calibration result by pressing the OK button.
9.8.9 3D Geometry Calibrations
These programs create the calibration data for reconstructing 3D images.
NOTICE! This calibration produces X-rays. Protect yourself from radiation.
Perform the calibrations in this order:
• 3D Geometry Calibration, 5x5 FOV
• 3D Geometry Calibration, 6x9 FOV
• 3D Geometry Calibration, 9x11 FOV
• 3D Geometry Calibration, 9x14 FOV
3D Geometry Calibration Phantom is required when performing these calibrations.
1. Attach the 3D Geometry Calibration Phantom to the device.
2. Select a 3D Geometry Calibration from the Calibrations menu.
3. Protect yourself from radiation.