KaVo ORTHOPANTOMOGRAPH OP 3D User And Installation Manual page 62

Orthopantomograph
Hide thumbs Also See for ORTHOPANTOMOGRAPH OP 3D:
Table of Contents

Advertisement

9 Installation
25. Lift the carriage onto the column using the installation rail as a hinge and by lifting
from the cutouts in the plywood support.
NOTICE! The liftable weight corresponds to approximately 55 kg and lifting it
on the column may require two persons.
NOTICE! Ensure that nothing gets caught between the carriage and the
column.
NOTICE! A latch locks the carriage in place when it is lifted on to the column.
26. Lift the carriage locking bolts (3 pcs) up and tighten them to 25 Nm torque to secure
the carriage.
CAUTION! The carriage may fall down during use, if the carriage locking
bolts are not properly installed and tightened.
27. Detach the plywood cover from one side first by detaching all the seven (7) indicated
screws and pulling it off.
62
ORTHOPANTOMOGRAPH
OP 3D

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents