X–ray generator
Generator
Nominal power
Max. electric output
Anode voltage
Anode current
Exposure time accuracy
DAP display accuracy
Supply frequency
Spine compensation
Accuracy of radiation output
Panoramic / 3D image detector
Technology / Sensor type
Image receptor area
User interface
Program and technical factors selection,
exposure control
Patient positioning
Device physical measures
Source–Image distance (SID)
Dimensions (H x W x D)
Weight
ORTHOPANTOMOGRAPH
™
OP 3D
High frequency DC generator
1520 W nominal @ 95 kV, 16 mA
1,7 kW
60 – 95 kV (+/– 5 kV)
2 – 16 mA (+/– 20%)
+/– 5%
+/– 25%
50 – 100 kHz
kV / mA compensated
> 95%
CMOS
147 x 112 mm / 5.78" x 4.4"
GUI, optional remote exposure switch,
exposure button with 3m cable.
Positioning panel, integrated
580 mm / 22.83"
1670–2467 x 765 x 1300 mm
65.7–97.1" x 30.1" x 51.1"
113 kg / 249 lbs