7 Maintenance
2. GUI: Select Quality Control menu.
3. GUI: Select 3D QC program from the list.
4. Attach the 3D QC phantom holder and the 3D QC phantom to the device.
5. Protect yourself from radiation.
6. Press and hold down the exposure switch to take the QC image.
7. The QC image preview appears on the GUI.
8. Acknowledge the result by pressing the OK button on the GUI.
9. The QC image appears on the workstation, showing the result of the check.
10. Workstation: Visually evaluate the image using the dental imaging software for
visual defects such as artifacts.
NOTICE! The device determines if the QC image is PASSED or FAILED
according to measured data, not based on the visible image quality.
NOTICE! You should also compare the new QC image to the reference image
taken during the installation or the latest service. Doing this helps you ensure
that the image quality has remained constant.