28. Detach the upper and rear screws (4 pcs) of the plywood cover on the other side and
carefully pull it off with the metal supports attached.
29. Detach the plywood top cover and the metal supports on top of the carriage (4
screws).
30. Remove the metal support from the rear of the carriage (4 screws).
31. Remove the protective plastic from the carriage carefully so that it doesn't catch and
break any circuit boards.
ORTHOPANTOMOGRAPH
NOTICE! Hold the plywood cover in place when detaching the last screw to
prevent it from falling down.
NOTICE! Be careful not to scratch the unit with the metal supports attached
to the plywood cover.
™
OP 3D
9 Installation
63