Available Accessories - Keysight InfiniiMax III+ Series User Manual

Probe
Table of Contents

Advertisement

1
Using InfiniiMax III+ Series Probes

Available Accessories

The following is a partial list of available probe and oscilloscope accessories.
Performance verification fixture
An E2655C or N5443A Performance Verification (PV) fixture is an
accessory that is used to properly position the probe during
performance verification testing as described in
Chapter
N5450B Extreme temperature cable extension kit
For extreme temperature testing, such as monitoring a
device in a temperature chamber, order the N5450B
InfiniiMax extreme temperature extension cable for use
with the N5441A solder-in head.
Keysight's Infiniimax probe amplifiers have a specified
operating temperature range from 5
N5441A probe head can withstand temperatures from –
55°C to +150°C for up to 250 test cycles. Use the extension
cables to physically separate the amplifier from the probe
head which allows you to operate the probe head inside a temperature chamber
while the probe amplifier remains outside the chamber.
None of the N2830/1/2A or N7000/1/2/3A probe amplifiers can withstand the
CAUTION
extreme temperatures (–55°C to +150°C) that the N5450B can withstand. When using
the N5450B extension cable, do not subject the InfiniiMax III+ probe amplifier to
extreme temperatures.
Besides the N5441A probe head, none of the other probe heads are designed for
CAUTION
extreme temperature testing. For more information about the extreme temperature
probing solution for Keysight oscilloscopes, check out the Keysight data sheet with
the literature number, 5990-3504EN.
To ensure a high-quality measurement, the N5450B cable set have been
phase-matched at the factory. A coupling tag is included with the cables to ensure
the cables stay as a matched pair. To install the coupling tag, slip the small end of
each cable through the holes in the tag. The tag can be positioned anywhere along
the length of the cable and can withstand the temperature ranges specified.
32
8.
Chapter 7
o
o
C to 40
C, but the
InfiniiMax III+ Series Probes User's Guide
and

Advertisement

Table of Contents
loading

This manual is also suitable for:

N2830aN2831aN7001aN7003aN7002aN7000a ... Show all

Table of Contents