Bandwidth Performance Verification - Keysight InfiniiMax III+ Series User Manual

Probe
Table of Contents

Advertisement

7
Performance Verification for N2830A-Series Probes

Bandwidth Performance Verification

This procedure documents the bandwidth performance of the N2832A InfiniiMax
III+ probe amplifier with the N2836A solder-in probe head.
The recommended test interval is one year/2000 hours.
N OT E
Equipment Needed
• InfiniiMax III+ N2836A solder-in probe head.
• Keysight 2 port E8361A/C Vector Network Analyzer or equivalent VNA that
This procedure is written assuming the E8361A/C PNA. If a different VNA is used, references
N OT E
that are specific to the PNA will need to be modified.
• Keysight N4692A-00F 2.92 mm (female/female) ECal module. Or, other
• Proper test port cables, with adapters as needed, to provide male 2.92 mm
• Keysight N5443A Performance Verification (PV) Fixture. The N5443A includes
• Maury Microwave 8775B2 2.92 mm (m) broadband load. Or other 2.92 mm (m)
• Keysight N5477A AutoProbe II to 3.5 mm (f) Adapter. The N5477A comes with
• Keysight N5442A AutoProbe I (BNC) to AutoProbe II Adapter.
• Keysight 1143A Probe Offset Control and Power Module.
• BNC 50 ohm male terminator. Or equivalent; not a critical part. For example, a
116
covers at least a 50 MHz to 13 GHz range. The VNA must have the following
capability:
Ability to use a Touchstone file to de-embed at a port.
A bias port for port 1 of the VNA. That is it must have an internal bias T's and
a BNC port that allows bias to be applied to port 1.
2.92 mm calibration kit that can calibrate to the 2.92 mm male connectors at
the test ports.
connectors at reference planes. If 2.4 mm or 1.85 mm test port cables are used,
the following Keysight adapters can be used to convert to 2.92 mm male
connectors:
11904A . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.4 mm (m) to 2.92 mm (m)
11904D. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.4 mm (f) to 2.92 mm (m)
an APC 3.5 (f-f) adapter (1250-1749).
load with similar or better return loss. A high quality 2.92 mm adapter to a
2.4 mm or 1.85 mm VNA calibration load with required return loss could be
used.
a required NMD 3.5 mm (m) to 3.5 mm (f) adapter (5062-1247).
Pomona number 3840-50 or 4119-50.
InfiniiMax III+ Series Probes User's Guide

Advertisement

Table of Contents
loading

This manual is also suitable for:

N2830aN2831aN7001aN7003aN7002aN7000a ... Show all

Table of Contents