Performance Verification For N7000-Series Probes - Keysight InfiniiMax III+ Series User Manual

Probe
Table of Contents

Advertisement

Keysight InfiniiMax III+ Series Probes
User's Guide
8 Performance Verification for
N7000-Series Probes
Bandwidth Performance Verification
138
DC Input Resistance Performance Verification
147
Performance Test Record
149
This chapter describes the equipment and procedures needed to verify the
performance of N7000/1/2/3A InfiniiMax III+ probes. The performance measured
in this chapter is of the probe by itself. Keysight high performance real-time
scopes (and sampling scopes under certain conditions) will apply probe correction
that will further enhance the performance of the probes.
Due to the very high frequency of the InfiniiMax III+ probing system, it is important to carefully
N OT E
adhere to the techniques and procedures described in this chapter to accurately measure the
performance.
Electrostatic discharge (ESD) can quickly and imperceptibly damage or destroy high
CAUTION
performance probes, resulting in costly repairs. Always wear a wrist strap when
handling probe components and insure that cables are discharged before being
connected.
137