Table 3-1 Status Screen Parameters (Continued)
Parameter
Syr
Vial
Ndl Wash
Sample
Column Temp
Injector
486 (1) and (2)
λ
Lamp
AUFS
2487 (1) and (2)
λ
Lamp
AUFS
Indicates the current position of the syringe. Valid positions are:
Home – Default position
Empty – Highest position (to purge the syringe)
Full – Lowest position (to draw in sample)
Indicates the number of the vial that is currently under the
injector.
Indicates whether the needle-wash pump is on.
Indicates the current pressure in the sample loop.
Indicates the desired and current temperature of the column (if
the column heater is installed).
Indicates the current position of the injector. Valid positions are:
Stream – In the high pressure stream
Seal – In the lower seal position
Wet – In the needle wash position
Vial – In the sample vial
Indicates the current status of the following parameters for the
486 detector(s) (if installed).
Wavelength
Lamp On or Off
Absorbance Units Full-Scale
Indicates the current status of the following parameters for the
2487 detector(s) (if installed):
Wavelength
Lamp On or Off
Absorbance Units Full-Scale
Description
Performing a Direct Control Run
33