To Enable The Auto Shutdown Function; Auto-Shutdown Dialog Box - Waters 2690 Quick Start Manual

Separations module
Table of Contents

Advertisement

• No changes sent to the Separations Module from a remotely connected Millennium
Chromatography Manager workstation, MassLynx computer, or external
autosampler
• An error condition that suspends the operation of the Separations Module
You can leave the Separations Module in the shutdown state indefinitely. The separation
method you specify in the Auto Shutdown dialog box defines the initial conditions that are
applied after the specified period of inactivity. Use the auto shutdown function:
• When there is a long delay between injections
• To minimize solvent flow after an unattended or long run
• To minimize sparge gas consumption
• To disable the vacuum degasser
• To disable temperature controls
• To turn off detector lamps

To enable the auto shutdown function:

1. Press the
Config
once.
2. Press the
Auto Shutdown
(Figure
2-5).
3. Select the separation method to be used while the Separations Module is shut
down. Only the initial conditions in the method you select are used.
screen key in the Main screen, then press the
screen key. The Auto-Shutdown dialog box appears
Figure 2-5 Auto-Shutdown Dialog Box
Preparing the Separations Module for Operation
More
screen key
15