If the "Ω Value Table" selection is used, touch the desired resistance value and to
select it. To exit the menu, touch the "x" on the top right of the table.
5. Push or if necessary.
6. Push .
7. Push . If a resistance value that is unavailable (as in the subsequent example
of 490 Ω) is entered, the display prompts to try again.
12/21/12
10:52am
Standby
External
External
Sense
Sense
OFF
OFF
8. After a valid resistance is specified and is pushed, the Calibrator clears the
entry from the entry bar near the bottom and shows in the larger area above.
9. Push . This resistance is now available at the output terminals.
3/15/2013
10:52am
Standby
1.9000096
External
External
Sense
Sense
OFF
OFF
+/- 0.00050mV (1 year) 213 days since calibration
Auto Range 220 mV
0.00000
Error 820: Calibrator Cannot Source
That Value.
External
Guard
OFF
+/- 10.0 ppm (1 year) 213 days since calibration
Expected Value 1.9 kΩ
42.00000
dBm
1.0000
kHz
External
2 Wire
Guard
Comp
OFF
ON
Front-Panel Operation
mV
OK
Wideband
Setup
Menu
OFF
kΩ
Wideband
Setup
Menu
OFF
4
Set the Output
hhp123.eps
hhp124.eps
4-25